Cleaning the sampling cone assembly
March 14, 2014, 715004387 Rev. B
111
Disassembling the sampling cone assembly
Required material
•
Chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
To disassemble the sampling cone assembly:
1. Slide the collar to the end of the combined 2.5-mm Allen wrench and
cone extraction tool.
Warning:
To avoid personal contamination with biohazards or
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the sampling cone
assembly.
Collar
Содержание Xevo TQD