4 Maintenance Procedures
160
March 14, 2014, 715004387 Rev. B
Replacing the IonSABRE II probe sample capillary
Replace the stainless steel sample capillary in the IonSABRE II probe if it
becomes blocked and you cannot clear it, or if it becomes contaminated or
damaged.
Removing the existing capillary
Required materials
•
Chemical-resistant, powder-free gloves
•
7-mm wrench
•
Combined 2.5-mm Allen wrench and cone extraction tool
To remove the existing capillary:
1. Remove the probe from the source (see
2. Retrieve the combined 2.5-mm Allen wrench and cone extraction tool
from its storage location on the source adaptor housing.
3. Use the combined 2.5-mm Allen wrench and cone extraction tool to
remove the 3 screws retaining the probe end-cover.
Warning:
To avoid personal contamination with biohazards or
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the probe and source
components.
Warning:
To avoid burn injuries, take great care while performing
this procedure; the probe and source can be hot.
End-cover retaining screws
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...