ESI mode
March 14, 2014, 715004387 Rev. B
59
3. With the probe label facing you, carefully slide the ESI probe into the
hole in the probe adjuster assembly, ensuring that the probe location
dowel aligns with the location hole of the probe adjuster assembly.
4. Tighten the probe locking ring to secure the probe in place.
Requirement:
Fully tighten the probe locking ring. Doing so ensures a
successful result on the automatic pressure test, which runs when the
probe is correctly seated, and the source enclosure door is closed.
5. Connect the ESI probe’s cable to the high-voltage connector.
TP03129
Location hole of the probe
adjuster assembly
Probe location dowel
Probe label
Probe locking ring