I/O signal connectors
March 14, 2014, 715004387 Rev. B
237
I/O signal connectors
The instrument’s rear panel includes two removable connectors that hold the
screw terminals for I/O signals. These connectors are keyed so that they can
receive a signal cable inserted only one way.
I/O signal connectors:
Warning:
To avoid electric shock, separate all electrical connections to
the rear panel from hazardous voltages by double or reinforced
insulation. Circuits of this type are classified as safety extra low voltage
(SELV). Examples of circuits that are typically SELV include contact
closure inputs and outputs for autosamplers, and UV, RI, and
fluorescence detector signal outputs for LC/MS systems. The electrical
connections on the rear panel of this mass spectrometer are all SELV.
To avoid electric shock and damage to the instrument, do not apply more
than,
• ±30 V d.c. to the Analog (Out) connection.
• 30 V d.c. to the Stop Flow (Out), Inject Start (In), Switch 2 (Out),
Switch 3 (Out), and Switch 4 (Out) connections.
1
2
3
4
5
6
7
8
9
10
+
−
+
−
Analog (Out)
Analog (Out)
Ground
Not Used
Not Used
Stop Flow (Out)
Stop Flow (Out)
Ground
Switch 2 (Out)
Switch 2 (Out)
Connector I
1
2
3
4
5
6
7
8
9
10
+
−
+
−
+
−
+
−
Inject Start (In)
Inject Start (In)
Ground
Event (In)
Event (In)
Switch 3 (Out)
Switch 3 (Out)
Ground
Switch 4 (Out)
Switch 4 (Out)
Connector II
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...