Hazards associated with removing an instrument from service
When you remove the instrument from use to repair or dispose of it, you must
decontaminate all of its vacuum areas. These are the areas in which you can
expect to encounter the highest levels of contamination:
•
Source interior
•
Waste tubing
•
Exhaust system
•
Rotary pump oil (where applicable)
The need to decontaminate other vacuum areas of the instrument depends on
the kinds of samples the instrument analyzed and their levels of
concentration. Do not dispose of the instrument or return it to Waters for
repair until the authority responsible for approving its removal from the
premises specifies the extent of decontamination required and the level of
residual contamination permissible. That authority must also prescribe the
method of decontamination to be used and the appropriate protection for
personnel undertaking the decontamination process.
You must handle items such as syringes, fused silica lines, and borosilicate
tips used to carry sample into the source area in accordance with laboratory
procedures for contaminated vessels and sharps. To avoid contamination by
carcinogens, toxic substances, or biohazards, you must wear
chemical-resistant gloves when handling or disposing of used oil.
Bottle placement prohibition
Warning:
To avoid personal contamination with biohazards,
toxic materials, or corrosive materials, wear
chemical-resistant gloves during all phases of instrument
decontamination.
Warning:
To avoid puncture injuries, handle syringes, fused
silica lines, and borosilicate tips with extreme care.
Prohibited:
To avoid injury from electric shock or fire, and to prevent
damage to the workstation and ancillary equipment, do not place objects
filled with liquid—such as solvent bottles—on these items, or expose
them to dripping or splashing liquids.
Содержание Xevo TQD
Страница 2: ...ii March 14 2014 715004387 Rev B...
Страница 22: ...xxii March 14 2014 715004387 Rev B...
Страница 56: ...2 Preparing for Operation 56 March 14 2014 715004387 Rev B...
Страница 200: ...4 Maintenance Procedures 200 March 14 2014 715004387 Rev B...
Страница 216: ...A Safety Advisories 216 March 14 2014 715004387 Rev B...
Страница 242: ...B External Connections 242 March 14 2014 715004387 Rev B...
Страница 246: ...C Materials of Construction and Compliant Solvents 246 March 14 2014 715004387 Rev B...