Installing AI 1310/AS 1310 Autosampler
Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer
82
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Starting-up
To start-up your AI 1310/AS 1310 sampling system
1. Plug in the Vdc power cable of the external portable power supply level VI efficiency into the jack
marked
24 Vdc
located on the sampling unit back side.
2. Connect the power cord of the external power supply to the mains outlet.
The AI 1310/AS 1310 sampling system will automatically run the self-testing routine during
which the following automatic checks and settings are carried out:
• Alignment between AI 1310/AS 1310 sampling system and injector
• Check of the turret travel
• Acknowledgment of the installed sample tray
• Calculation of the syringe zero
Note
The self-test routine is automatically carried out every time the safety door of the
turret is closed.