Running the Flash IRMS as Stand-alone Instrument
Configuring the Analyzer
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
145
f.
In Main Menu for IRMS select
File | Instrument name and configuration
and click
Elemental analyzer setup
for visualizing the dialog window of
, where the
configuration parameters must be set.
Figure 152.
Configuration Dialog Window
g. In the field
Elemental Analyzer Connection
select the computer serial port (COM1, COM2,
and so on), to which the instrument is connected.
h. In the field
Instrument
Settings
choose the following settings:
• Line Frequency = 50 Hz
• TCD Settings Source = Internal
• TCD Settings Polarity = Positive
i.
In the field
Sampler Setting
select the type of autosampler installed on the instrument.
• In the case of autosampler for liquid samples, also specify the serial port of the computer
to which the autosampler is connected, and the number of vials.
• Click
Ok
, then click
Ok
to return to Main Menu
for IRMS.
2. In Main Menu for IRMS click the icon
. The following window appears where the analyzer
operating parameters are visualized. See
Table 29.
Main Menu for IRMS: Description of Icons (Graphic Buttons)
Icon
Function
Description
Edit elemental analyzer
parameters
Gives the access to the three pages containing the commands
for setting temperatures, flows, times, detector, and the
analyzer control functions.
Elemental analyzer status
Comprises four pages displaying the analyzer conditions.