![Thermo Scientific EA IsoLink IRMS System Скачать руководство пользователя страница 63](http://html.mh-extra.com/html/thermo-scientific/ea-isolink-irms-system/ea-isolink-irms-system_operating-manual_1105021063.webp)
2
Analytical Principles
Pneumatic Circuit
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
35
The external pneumatic connection is shown in
Figure 32.
External Pneumatic Connection for NCS and OH Analysis
Pressure Regulators
The pressure regulators allow the manual adjustment of the carrier gas (helium or argon) and oxygen
inlet pressures. The regulators are located in the detector compartment and they are schematically
shown in
.
Figure 33.
Pressure Regulators
The pressure regulators are common in all the configurations of the EA IsoLink IRMS System for
CNSOH. The regulators consist of the following components. See
NCS Configuration uses the
right
channel.
Note:
The reference flow enters the
pyrolysis reactor from the bottom
before it ends as purge in the
Autosampler NCS.
OH Configuration uses the
left
channel.
Note:
The carrier gas flow is not
shown here. It enters the pyrolysis
reactor from the bottom.
The reference flow enters the system
at the combustion reactor and ends
as purge for the Autosampler H/O.
1
2
1
2
To ConFlo
(Purge Autosampler H/O)
1
2
3
4
(Purge Autosampler NCS)
CARRIER
(Carrier Gas NCS)
REFERENCE
OH
NCS
Cap
1
2
1
2
To ConFlo
(Purge Autosampler H/O)
1
2
3
4
(Purge Autosampler NCS)
REFERENCE
OH
NCS
Cap
Table 13.
Pressure Regulators Pneumatic Diagram (Sheet 1 of 2)
Component
Description and function
He
Helium inlet port
O
2
Oxygen inlet port
PRV1
Helium pressure regulator
PI1
Helium
pressure gauge
He
O
2
PI1
PI2
PRV2
PRV1