9
Running the Flash IRMS as Stand-alone Instrument
Using Isodat Software Suite and EagerSmart Data Handling Software at the Same Time
152
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Using Isodat Software Suite and Eager
Smart
Data Handling Software
at the Same Time
To install the software version
1. Make a backup of Isodat Software Suite with the Version Handler if Isodat Software Suite is
installed.
2. Make an additional backup of the
User
folder under Isodat Software Suite.
3. Close down Isodat Software Suite completely by right-click task bar.
4. Install Eager
Smart
Data Handling Software
.
To switch fromEager
Smart
Data Handling Software to Isodat Software Suite
1. Before using either software make sure the other software is completely closed.
2. Open
Acquisition
. The OCX is automatically registered for Isodat Software Suite if Eager
Smart
Data Handling Software is completely closed down first. If problems occur a Safety Cut-Off
message is visualized. Go to the Isodat Software Suite
Configurator
and
Re-Register OCX
, then
wait 10 seconds.
3. Close
Configurator
and start
Acquisition
.
To switch from Isodat Software Suite to Eager
Smart
Data Handling Software
1. Before using either software make sure the other software is completely closed.
2. Start Eager
Smart
Data Handling Software. The OCX is automatically registered for Eager
Smart
Data Handling Software if Isodat Software Suite is completely closed down. If problems occur you
may register and unregister the OCX using the two bat –files in the following folder of Eager
Smart