1
Getting Familiar with your EA IsoLink IRMS System for CNSOH
Instrument Basics
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
3
* Required by Isodat 3.0
Instrument Basics
The Flash IRMS Elemental Analyzer is part of the EA IsoLink CNSOH IRMS System.
The Elemental Analyzer comprises:
•
Two furnaces
— The
left furnace
, for high temperature conversion analysis, contains a silicon
carbide heater element. The heater element is immersed into a refractory material contained in a
metallic compartment. This furnace is used only for the
H
and
O
isotope ratio determinations.
The
right furnace
contains a quartz candle on which an electric resistance is wound.
The candle is immersed into a refractory material contained in a metallic compartment.
This furnace is used only for the
N
,
C
and
S
isotope ratio determination.
•
Furnace Temperature Regulation
— The temperature is monitored by a thermocouple of
Pt-Pt/Rh type appropriately located in the furnace.
•
Furnace Cooling
— The cooling time varies according to the operating temperature setting and
room temperature.
•
Thermal conductivity detector (TCD)
— Located in a thermostatic chamber at controlled
programmable temperature. This chamber also accommodates the analytical column.
•
Chromatographic columns
— The chromatographic column performs the chromatographic
separation of the reaction products generated during the combustion or high temperature
conversion process. The chromatographic column for high temperature conversion is placed in the
thermostatic chamber of the TCD detector according to the instrument configuration.
The analytical column for the combustion side (N, C and S analysis) is placed in a separate ramped
GC oven which can be heated and cooled in a fast way.
Table 3.
Minimum Requirements for Personal Computer
Components
Description
Computer
Any PC can be used, including laptop computer
Operating System: Windows®
2000 / XP* / Vista / 7* / 8
Pentium Processor minimum 256 MHz
Hard drive with at least 1 GB free
One free COM port for instrument control
One free COM port for balance, if required
One free USB port
Monitor
Color 1024 x 768 or better
Printer
Any printer accepted by the operating system
Description
Qty
Part number
H/O separation column (SS; 1 m; 1/8-in. OD)
1
260 08240
IRMS separation column (SS; 60 cm; 1/8-in. OD)
1
260 08241
IRMS NC separation column (SS; 60 cm; 1/8-in. OD)
1
260 08242
Sulfur separation column (Teflon®; 80 cm; 1/8-in. OD)
1
260 08243