6
Using EA IsoLink IRMS System for CNSOH
Starting a Sulfur Measurement
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
117
2. First
SO
2
Reference Gas
pulse activated at
20 s
(duration:
20 s
).
3. Second
SO
2
Reference Gas
pulse activated at
90 s
(duration:
20 s
). It is assigned as Standard pulse
for
δ
value calculation. See
Time
column in
4. EA starts at
105 s
. At this time the oxygen is injected into the reactor.
5. Autosampler activated by EA after a sampling delay defined in the EA method, e.g. 10 s after
oxygen injection.
6. Sample peak appears approximately
100 s
after Autosampler activation.
7. Acquisition stops at
400 s
.