9
Running the Flash IRMS as Stand-alone Instrument
Configuring the Analyzer
144
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Figure 150.
Instrument Name and Configuration
a.
In the field
Instrument name
, type the instrument
serial number
(6 digits; for example
991234). See the label located on the instrument rear panel.
b. In the field Analytical Instrument Configuration select Undefined.
c.
In the field
IRMS & Argon
select
Instrument control for Flash EA IRMS (NC) and HT
.
d. Click
OK
. The system ask to exit and reboot Eager
Smart
.
e.
At the reboot of Eager
Smart
, the simplified Main Menu for IRMS is visualized. See
.
Figure 151.
Main Menu for IRMS
Menus and buttons of the Main Menu for IRMS are described in the following
and
respectively.
Table 28.
Main Menu for IRMS: Description of Menus
Menu
Description
Sub-menus and Options
File
Monitors the analysis in real time.
• Instrument name and configuration
• View sample being acquired
• Exit
Eager
Smart
Tools
Used when the ashes removal, the
reactor replacement, or both, are
required as maintenance.
• Ashes removal
• Reactor replacement
Help
Accesses to the help system of
Eager
Smart
.
• About
Eager
Smart