6
Using EA IsoLink IRMS System for CNSOH
Creating an EA Method for H+O Measurement
106
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Figure 120.
Defining Results Export and Printout Parameters
2. Select your destination
Folder Name
and
File Name
for the
Results
.
3. Select the
Format
for the export file.
4. Select
Printout,
for example, Yes.
5. Click
OK
.
Dual Measurement Results
The following events occur during the acquisition:
• Peak center procedure is performed.
• The H
2
reference gas pulse is activated after 20 seconds for 20 seconds.
• The H
2
reference gas pulse is activated after 60 seconds for 20 seconds.
• The sample is dropped into the reactor.
• The H
2
sample peak appears approximately 15 seconds after reaction start. See
.
• The CO sample peak appears approximately 100 seconds after reaction start. See
.
• The CO reference gas pulse is activated after 340 seconds for20 seconds.
• The CO reference gas pulse is activated after 400 seconds for 20 seconds.
• The acquisition stops after 450 seconds.
• After finishing data acquisition, the printer creates a data output sheet as defined by the Result
Workshop template (
*.irw
) selected. The results are exported to a spreadsheet file if defined in
.