![Thermo Scientific EA IsoLink IRMS System Скачать руководство пользователя страница 65](http://html.mh-extra.com/html/thermo-scientific/ea-isolink-irms-system/ea-isolink-irms-system_operating-manual_1105021065.webp)
2
Analytical Principles
Pneumatic Circuit
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
37
Automatic Switching Valve Block (ASV)
Allows the automatic switching from the
combustion
N,C, S configurations to the
pyrolysis
O, H
configuration, and vice-versa. The module is schematically shown in
, and consists of the
following components. See
.
Figure 35.
Automatic Switching Valve Bock Pneumatic Diagram
S2
Electronic flow sensor for helium
as reference gas. It cooperates with the EVP2 electronic
controller (proportional valve).
EVP1
Electronic flow controller for helium as carrier gas and oxygen to control the flow rates of gases
according to the flow values set.
EVP2
Electronic flow controller for helium as reference gas to control the flow rate according to the
required flow value.
Table 14.
Parts of the EFC-t Module (Sheet 2 of 2)
Component
Description and function
2
1
EV1
(2)
(1)
(7)
(6)
(4)
(5)
(8)
(9)
(3)
HO
NCS
MS
Purge HO
2
1
EV2
2
1
EV3
2
1
EV4
Purge NCS
Carrier Gas
Reference Gas
(2)
(1)
(7)
(6)
(4)
(5)
Purge NCS
(8)
(9)
AUTOMATIC
SWITCHING BLOCK
(3)
HO
NCS
MS
Purge HO
HIGH TEMPERATURE MODE
PYROLYSIS, O, H
3
3
3
(2)
(1)
(7)
(6)
(4)
(5)
Purge NCS
(8)
(9)
(3)
HO
NCS
MS
Purge HO
COMBUSTION NCS
AUTOMATIC
SWITCHING BLOCK
EFC-t
ASV