2
Analytical Principles
The Ramped GC Oven
50
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Figure 48.
Nuts and Seals for Adsorption Filter
4. Pour sequentially the required filling materials into the adsorption filter, ensuring that each layer
has the indicated size. At each step gently press the quartz wool using the rod provided.
5. Do the last layer using a sufficient quantity of quartz wool to form the required layer.
6. Complete the procedure by screwing on the second nut complete with its seal, as shown in
.
Figure 49.
Preparation of the Adsorption Filter
The Ramped GC Oven
The Ramped GC Oven is controlled by a JUMO controller. The controller is
pre-programmed and locked. It allows to switch between two setpoints,
SP1
and
SP2
.
•
SP1
is the start temperature of 70 °C
•
SP2
is the end temperature of 240 °C
When receiving the GC start signal (GC On) from Isodat, the Ramped GC Oven increases
automatically the temperature from SP1 to SP2. When stopping the GC (GC Off ) the oven cools
down to SP1. The starting temperature of 70 °C is defined to separate the N
2
and CO
2
peaks.
After the CO
2
peak a rapid increase of the temperature accelerates the elution of the SO
2
.
Detailed information can be found in the
Ramped GC Oven Operating Manual
.
Seal
Nut