Using EA IsoLink IRMS System for CNSOH
Setting Up a HO Method
100
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
14. Set the
Evaluation@H2
parameters for your new dual measurement method as shown in
.
15. Click the tab
Peak Detection@H2
, see
Figure 109.
Peak Detection@H2 Page of the Dual Measurement Method
16. Set the
Peak Detection@H2
parameters for your new dual measurement method as shown in
.
17. Click
Advanced Parameter
to open the advanced parameter settings, see
Figure 110.
Peak Detection@H2 Page – Advanced Parameters of the Dual Measurement Method
18. Set the
Advanced Parameters
parameters for your dual measurement method as shown in
.
19. Select the
Printout@H2 tab
of your new dual measurement method, see
.
Figure 111.
Printout@H2 of the Dual Measurement Method
Note
The values for reference gas and Amount% are examples and must be adjusted according
to the isotope value of the reference gas and amount% of H in your sample. Please refer also to
the
ConFlo IV manual, Chapter 4
, where the settings for reference/blank are described.