Installing AI 1310/AS 1310 Autosampler
Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer
78
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Figure 78.
Remove Protection
3. Insert the centering plate into its seat located in the
right
side section of the sampling unit base
paying attention that the guide hole, present on the arm of the centering plate, correctly fits the
injector nut. See
Figure 79.
Centering Plate
.
Protection
Injection Nut
Centering Plate