5
Installing AI 1310/AS 1310 Autosampler
Installing the Sampler Support on the Flash IRMS Elemental Analyzer
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
75
Figure 73.
Mounting Support Bracket
2. Mount the sampler support. See
.
Figure 74.
Mounting Sampler Support
a.
Insert the provided fixing screw into each slot present on the support.
b. Insert each screw into the relevant spacer paying attention to keep its largest surface turned
toward the support base.
c.
Hold the spacers in position with their flat side toward the inside, then place the sampler
support on the support bracket.
d. Guide the two fixing screws located on the external spacers into the corresponding fixing
holes.
e.
Loosely tighten the screws.