Using EA IsoLink IRMS System for CNSOH
Create a Gas Configuration for a Sulfur Measurement
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
113
18. Type
No
. If you would type Yes, this would automatically mark the template (i.e.
CO
2
) instead of
SO
2
in the
Ratio Groups
window below.
a.
Mark
SO
2
.
b. If
Ratio Groups
other than
SO
2
are marked, clear them all.
19. Confirm by
OK
.
20. The new Gas Configuration
SO
2
appears in the list as a row of its own.
Figure 125.
Creating a New Gas Configuration
21. In the
Calibration
column select your current calibration file.
shows a common cup configuration as used in most Delta mass spectrometers,
i.e. universal triple collector. If you have a special cup configuration, the respective masses will be
collected in other cups.