1
Getting Familiar with your EA IsoLink IRMS System for CNSOH
Furnaces Compartment
14
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
The furnaces are accessible by removing the protecting plate. See
Figure 13.
Furnaces Compartment
In the software, the furnaces are limited to the following maximum temperatures to avoid reduced
lifetime and damage:
• LEFT Furnace: 1450 °C
• RIGHT Furnace: 1100 °C
The furnace temperature is monitored by a thermocouple located inside the furnace. The furnaces are
cooled when required by the operator or if there is a safety cut-off. The cooling time depends on the
operating temperature and the room temperature.
Protection Plate
Left Furnace for high temperature conversion analysis (OH)
Right Furnace for combustion analysis (NCS)