1
Getting Familiar with your EA IsoLink IRMS System for CNSOH
Electrical Compartment
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
19
Figure 18.
Electrical Compartment Internal View
Low Voltage Compartment
It contains the electronic boards to operate and control the instrument. These boards are interlocked
through a mother board.
reports the function of each electronic board present in the low voltage section:
Table 4.
Description of the Function of the Electronic Boards (Sheet 1 of 2)
Board
Function
MB 1112
Mother board. It provides interlocking between low voltage boards and with
the rest of the instrument. This board can be connected to a NiCd 3,6 V; 280
mA/h rechargeable battery located nearby. The rechargeable battery
replacement must be performed by specialized technical personnel
CPU 1112
This board has full control of the instrument operation.
It controls the communication between operator and machine through
Eager
Smart
Data Handling Software. Actuates the Safety Cut Off device,
which puts the instrument in safe conditions, when an alarm condition occurs.
HWD 1112
Provides power supply to the TCD detector filaments. Allows the detector
oven thermo-regulation and also amplifies and converts the detector signal to
send it to the PC.