Installing MAS Plus Autosampler
MAS Plus Autosampler Installation
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
69
6. If additional sample trays are required, install them in the correct order one over the other, paying
attention that the relevant locating marks are in correspondence with the relevant seats marked
0
(zero), and placing the samples properly.
IMPORTANT
Before installing an additional sample tray, make sure that the samples to analyze are
placed in all the seats of the previous tray installed.
CAUTION
Before starting samples analyses, make sure that the protection cover is positioned over
the sample tray. A complete deaeration of the area where samples are housed is only possible if the
cover is in place.
Pay attention not to invert the cover; the surface marked “
Side-up
” must be turned towards you.