7
Maintenance
Current Maintenance Program
128
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
2. Undo the 4 fixing screws securing the protection panel. See
Figure 138.
Access to the Detector
shows the detector compartment, the heating block where the detector is housed, and
the gas chromatographic column.
Figure 139.
Compartment Housing the Chromatographic Column
3. Unscrew the fittings from the column ends and remove the column from the compartment.
4. Introduce the new column and connect its ends to the fittings.
5. Re-mount the protection panel and the adsorption filters.
Note
According to the instrument configuration there can be one or two analytical columns.
Securing Clips for
Adsorption Filters
Fixing Screws
Protection
Plate
Heating Block
TCD Detector
Chromatographic Column