5
Installing AI 1310/AS 1310 Autosampler
This chapter provides the instruction for installing the AI 1310/AS 1310 autosampler for liquid
samples on the Flash IRMS Elemental Analyzer.
Preliminary Information
This section contains the preliminary information for the installation and the connection of the
AI 1310/AS 1310 sampling system to the Flash IRMS Elemental Analyzer.
Who Performs the Installation
The AI 1310/AS 1310 is installed by authorized Thermo Fisher Scientific technical engineers, who will
check its correct operation. For more details, please contact Thermo Fisher Scientific local
representatives. Should the instrument not be installed by Thermo Fisher Scientific personnel, strictly
adhere to the instructions reported in this section.
Electrical Requirement
The instrument has the following power supply rating:
• 24 Vdc through a portable external power supply, level VI efficiency
−
input 100-240 Vac; 50-60 Hz — output 24 Vdc; 3 A - 3.75 A
Contents
•
•
Installing the Direct Injection Device for Flash IRMS Elemental Analyzer
•
Installing the Sampler Support on the Flash IRMS Elemental Analyzer
•
Installing the AI 1310/AS 1310 on the Flash IRMS Elemental Analyzer
WARNING
YOU MUST ONLY USE THE PORTABLE EXTERNAL POWER SUPPLY
FURNISHED WITH THE INSTRUMENT BY THERMO FISHER SCIENTIFIC.