2
Analytical Principles
Pneumatic Circuit
38
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Table 15.
Parts of the Automatic Switching Block
Component
Description and function
EV1
In combustion NCS analysis controls the reference gas flow through the right channel.
In pyrolysis O,H analysis controls the carrier gas flow through the right channel.
EV2
In combustion NCS analysis controls the carrier gas flow through the right channel.
In pyrolysis O,H analysis controls the reference gas flow through the right channel.
EV3
In combustion NCS analysis controls the carrier gas flow to the ConFlow.
In pyrolysis O,H analysis controls the reference gas flow to the ConFlow.
EV4
In combustion NCS analysis controls the purge gas flow to the right autosampler.
In pyrolysis O,H analysis controls the purge gas flow to the left autosampler.
Carrier Gas
Reference Gas
He
O
2
PI1
PI2
PRV2
PRV1
He
O
2
F
F
EV1
6.5 mL/min.
300 kPa
3
2
1
2
1
S1
EVP1
EV2
2
3
1
2
1
EVP2
S2
F
F
1
EV3
2
EV4
1
2
(2)
(1)
(7)
(6)
(4)
(5)
(2)
(1)
(7)
(6)
(4)
(5)
Purge NCS
(8)
(9)
EFC - T
AUTOMATIC
SWITCHING BLOCK
(3)
HO
NCS
MS
Purge HO
Carrier Gas
Reference Gas
He
O
2
PI1
PI2
PRV2
PRV1
He
O
2
F
F
EV1
6.5 mL/min.
300 kPa
3
2
1
2
1
S1
EVP1
EV2
2
3
1
2
1
EVP2
S2
F
F
1
EV3
2
EV4
1
2
(2)
(1)
(7)
(6)
(4)
(5)
(2)
(1)
(7)
(6)
(4)
(5)
Purge NCS
(8)
(9)
EFC - T
AUTOMATIC
SWITCHING BLOCK
(3)
HO
NCS
MS
Purge HO
COMBUSTION NCS
HIGH TEMPERATURE MODE
PYROLYSIS, O, H