Analytical Principles
Preparation of Reactors and Adsorption Filter
44
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Preparation of Reactors and Adsorption Filter
NC Configuration
report the characteristics of the components required for NC determination, and the type and
size of the filling materials to be used for a proper preparation of the reactor.
shows the size of the filling material.
Figure 40.
Size of Filling Material
•
Reactors
— The reactor is a quartz tube having a conical bottom end.
There are two different sizes
available:
−
quartz glass reactor of 18 mm OD
Table 17.
Components required for NC Determination
Reference
Component
Characteristic
Filling Materials
Right Channel
Reactor
Material:
Quartz
Quartz Wool
Chromium Oxide
Reduced Copper
Silvered Cobaltous-Cobaltic Oxide
F2
Absorption Filter
Material:
Plexiglas
Quartz Wool
Magnesium Perchlorate (Anhydrone)
CC2
Gas chromatographic
Column
Material:
Stainless
Steel
---
Note
In some cases, it can be advantageous to use
copper oxide
instead of
chromium oxide
.
Reactors filled with copper oxide must be operated at 900°C - 920°C. Check your calibration with
appropriate standards to maintain analytical precision and quality after changing the reactor filling
from chromium oxide to copper oxide.
NC Configuration with EA IsoLink CN/OH
CC2
F2 Adsorption Filter: PlexiglasTube
10 mm
Quartz Wool
Quartz Wool
F2
10 mm
Magnesium Perchlorate
(Anhydrone)
TCD
Gas Chromatography Column
ConFlow IV
ConFlow IV
or “CF IV”
or “CF IV”
IRMS
IRMS
100 mm
50 mm
50 mm
20 mm
10 mm
Right Channel Reactor : Quartz Tube
Quartz Wool
Chromium Oxide
Copper Reduced
Quartz Wool
10 mm
Quartz Wool
10 mm
Quartz Wool
Silvered Cobaltous-Cobaltic Oxide
Copper Oxide
Right
Channel