Installing Flash IRMS Elemental Analyzer
Installing the Reactors into the Furnaces
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
57
Figure 51.
Protection Cap Removal
4. Remove the MAS Plus autosampler, if installed, by manually undoing its fixing nut. See
.
Protection Cap
Reactor Retaining
Nut
Washer
Conical O-ring
Coupling Union
Right Channel - Combustion)
Left Channel - Pyrolysis
Protection Cap
Reactor Retaining
Nut
Washer
Conical O-ring
Coupling Union
Bottom Feed
Connector
ATTENTION
Do not dispose of the protection cap but keep it in your lab for future use.
The protection cap should be reinstalled on the coupling union in case the instrument is not used
for a long time
. The washer and the conical O-ring removed are used for the installation of the
reactor.