Getting Familiar with your EA IsoLink IRMS System for CNSOH
Connections Panel
22
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
• A 15-pin connector marked
He
M
for the control of helium management device.
Power Supply Area
The
power supply
area comprises the following components. See
.
Figure 21.
Power Supply Are
• Breaker marked
Mains
to power the instrument ON/OFF.
−
Position
I
= instrument powered ON; — Position
O
= instrument powered OFF.
• 230 V; 50/60 Hz mains connector.
Gases Supply Area
The
gas supply
area comprises the gas inlet ports. See
Figure 22.
Gases Supply Area
• The
helium
and
oxygen
gas inlet ports, marked
He
and
O
2
are directly connected to the pressure
regulators.
details the pressure value to be set for each gas inlet port.
Gas pressures must be set and controlled through the pressure regulators and the pressure gauges of the
instrument.
provides indications on the most currently used units of pressure.
Table 7.
Gas Inlet Ports and Pressure Setting
Port
Description
Set TO:
He
Inlet port for helium or argon
250 kPa (2.5 bar, 36 psig)
O
2
Inlet port for oxygen
250-300 kPa (2.5-3 bar, 36-44 psig)
Helium Inlet Port
Oxygen Inlet Port