2
Analytical Principles
Pneumatic Circuit
36
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Electronic Flow Controller (EFC-t) Module
The EFC-t module is common in all the configurations of the EA IsoLink IRMS System for CNSOH.
The module is schematically shown in
, and consists of the following components.
See
Figure 34.
Thermo-regulated EFC-t Module Pneumatic Diagram
PRV2
Oxygen pressure regulator
PI2
Oxygen pressure gauge
Table 13.
Pressure Regulators Pneumatic Diagram (Sheet 2 of 2)
Component
Description and function
Table 14.
Parts of the EFC-t Module (Sheet 1 of 2)
Component
Description and function
He
Helium inlet port
O
2
Oxygen inlet port
F
Stainless steel filter
EV1
Two-way solenoid valve to control oxygen inlet.
EV2
Three-way solenoid valve to control helium inlet and to allow switching between helium
and
oxygen.
EV3
Two-way solenoid valve, normally open, to control the inlet of helium flowing back from the TCD
detector analytical channel. The valve is closed during the leak test.
EV4
Two-way solenoid valve, normally open, to control the inlet of helium flowing back from the TCD
detector reference channel. The gas leads to port 1 of the MAS Plus autosampler and then to the
ConFlo..
S1
Electronic flow sensor for helium as carrier gas and oxygen during the sampling stage.
It cooperates with the EVP1 electronic controller (proportional valve).
He
O
2
F
F
EV1
6.5 mL/min.
300 kPa
3
2
1
2
1
S1
EVP1
EV2
2
3
1
EFC - T
2
1
EVP2
S2
F
F
1
EV3
2
EV4
1
2
(2)
(1)
(7)
(6)
(4)
(5)
EFC-t
ASV