Troubleshooting
Safety Cutoff
138
EA IsoLink IRMS System for CNSOH Operating Manual
Thermo Scientific
Safety Cutoff
Instrument malfunctioning, due to a component failure or to abnormal operating conditions, is
identified by the red lighting of the Safety Cutoff LED indicator. See
Figure 145.
Safety Cutoff LED
When lit, this LED indicates that the furnaces and detector oven power has been cut off for safety
reasons.
The Safety Cutoff status is followed by an error message about the possible cause of error.
Peak between nitrogen and carbon
peaks.
Oxygen line contaminated.
Exclude autosampler and check the
oxygen blank.
Inadequate oxygen purity.
Use Oxygen with adequate purity.
Exclude autosampler and check the
oxygen blank.
Inadequate oxygen dosage.
Increase oxygen injection time.
High carbon blank.
Tin containers contaminated.
Check the tin container box, tweezers,
work bench are clean.
Memory effect due to bad combustion
of previously run analyses.
Remove ash and analyze lower
amounts of sample.
Decreasing nitrogen blank values.
Oxygen line contaminated.
Wait 10-20 minutes for complete
purging of the oxygen line.
Repeat blank analysis.
Increasing nitrogen blank values.
Copper exhausted
Replace the reactor.
Retention times very delayed respect
the normal chromatogram.
Presence of leaks in the pneumatic
circuit.
Perform Leak Test.
Presence of obstructions in the
pneumatic circuit.
Reach and remove the obstruction
dissecting the pneumatic circuit.
Table 25.
Analytical Troubleshooting Guide (Sheet 2 of 2)
Problem
Diagnosis
Remedy
Safety Cutoff LED