Using EA IsoLink IRMS System for CNSOH
H3-Factor Determination
Thermo Scientific
EA IsoLink IRMS System for CNSOH Operating Manual
107
• The chromatogram and result grid are shown in
,
and
:
Figure 121.
Dual Measurement Chromatogram
The H
2
page displays information about the hydrogen-related peaks, see
Peaks Nr.
1, 2 and 3 in
Figure 122.
Dual Measurement H
2
Result Grid
The CO page displays information about the CO-related peaks. See
Peaks Nr.
4, 5 and 6 in
.
Figure 123.
Dual Measurement CO Result Grid
H3-Factor Determination
Protonation reactions in the ion source result in H3+-ion production. The H3+ portion of the
m/z
ion
beam is determined as the H3-factor. The H3-factor is used to correct the H3+ contribution to the
m/z3 signal. A low and stable H3-factor is needed for a good DH/H2 determination.
H3-factor determination is based on the peak area information of a “standard gas on/off
chromatogram” (*.dxf ) with standard gas pulses of different amplitudes. How to perform a H3-Factor
Determination is explained in Chapter 3 of the
ConFlo IV Operating Manual
.