Operating Instructions
Model 400E Ozone Analyzer
Instruction Manual
92
P/N 04316 Rev: B
To adjust the zero and span values of the current outputs, activate the
ANALOG
I/O CONFIGURATION
MENU
, then press
:
DIAG AIO
CONC_OUT_2 SPAN: 10000 mV
U100 UP10
UP DOWN DN10
D100
ENTR EXIT
DIAG AIO
AUTO CALIBRATING CONC_OUT_2
DIAG AIO
AIN CALIBRATED:
NO
SET>
EDIT
EXIT
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO
CONC_OUT_2 RANGE: CURR
<SET
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 CALIBRATED: NO
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_2:CURR, NO CAL
< SET SET>
EDIT
EXIT
Press
SET>
to select the analog output channel
to be configured:.
Increase or decrease the current
output by 100, 10 or 1 counts.
The resulting change in output
voltage is displayed in the upper
line.
Continue adjustments until the
correct current is measured with
the current meter.
DIAG AIO
CONC_OUT_2 ZERO: 27 mV
U100 UP10
UP DOWN DN10 D100
ENTR EXIT
EXIT
ignores the
new setting,
ENTR
accepts the new
setting.
DIAG AIO
CONC_OUT_2 CALIBRATED: YES
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_2 ZERO: 9731 mV
U100 UP10
UP DOWN DN10 D100
ENTR EXIT
DIAG AIO
CONC_OUT_2 CURR-Z: 0 mV
U100
UP10
UP DOWN DN10 D100
ENTR EXIT
The instrument attempt to automatically calibrate
the channel … then beep.
If a current meter is not available, an alternative method for calibrating the current
loop outputs is to connect a 250
Ω
±
1% resistor across the current loop output.
Using a voltmeter, connected across the resistor, follow the procedure above but
adjust the output to the following values:
Table 6-13: Current Loop Output Calibration with Resistor
FULL
SCALE
VOLTAGE FOR 2-20
MA
(measured across
resistor)
VOLTAGE FOR 4-20 MA
(measured across resistor)
0%
0.5 V
1.0 V
100%
5.0 V
5.0 V
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...