Operating Instructions
Model 400E Ozone Analyzer
Instruction Manual
70
P/N 04316 Rev: B
6.4.2.
Setting the O
3
Generator Low-Span (Mid Point) Output
Level
For applications where To set the ozone LO SPAN (Midpoint) concentration for the
IZS O
3
generator, press:
EXIT
ignores the
new setting.
ENTR
accepts the
new setting.
SAMPLE RANGE = 500.0 PPB O3 =XXX.X
< TST CAL CALZ
CALS
SETUP
SETUP X.X 0) DAS_HOLD_OFF=15.0 Minutes
NEXT
JUMP
EDIT PRNT
EXIT
SPAN CAL M
3) O3_GEN_LOW1=
100.0 PPB
0 1 0 0 .0 ENTR
EXIT
Toggle these keys to change setting
•
Only Values from 0 to 1500 will be accepted..
•
A value of 0 turns the lamp OFF.
•
The
ENTR
key will disappear if an invalid setting is
attempted.
SETUP X.X
CFG ACAL DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
COMM
VARS
DIAG O3 HALT
SETUP X.X
ENTER VARS PASS:818
8 1 8
ENTR EXIT
Repeat pressing
NEXT
until …
SETUP X.X
3) O3_GEN_LOW1=
100.0 PPB
NEXT JUMP
EDIT
PRNT
EXIT
Sets LOW SPAN Point for
RANGE1.
To Set the LOW SPAN point
for RANGE2 in
DUAL
or
AUTO
range modes …
Press
NEXT
key once more
to select
O3_GEN_LOW2
then continue as shown.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...