Model 400E Ozone Analyzer
Instruction Manual
Appendix C
Warranty/Repair Questionnaire
Model 400E
TELEDYNE API CUSTOMER SERVICE
EMAIL:
PHONE: (858) 657- 9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816
04404 Rev B
1
TELEDYNE
INSTRUMENTS
Advanced Pollution Instrumentation
A Teledyne Technologies Company
CUSTOMER: ____________________________________________ PHONE: _______________________________________
CONTACT NAME: _______________________________________ FAX NO. _______________________________________
SITE ADDRESS: _________________________________________________________________________________________
MODEL 400E SERIAL NO.: ___________________ FIRMWARE REVISION: ______________________________________
1. ARE THERE ANY FAILURE MESSAGES? _______________________________________________________________
________________________________________________________________________________________________________
________________________________________________________________________________________________________
PLEASE COMPLETE THE FOLLOWING TABLE: (
NOTE
:
DEPENDING ON OPTIONS INSTALLED, NOT ALL TESTS
PARAMETERS BELOW WILL BE AVAILABLE IN YOUR INSTRUMENT
)
2. *IF OPTIONS ARE INSTALLED
Parameter
Recorded Value
Acceptable Value
RANGE
PPB/PPM
1 – 10,000 PPB
STABIL
<= 0.3 PPM WITH ZERO AIR
O3 MEAS
mV
2500 – 4800 mV
O3 REF
mV
2500 – 4800 mV
O3 GEN*
mV
80 mV. – 5000 mV.
O3 DRIVE*
mV
0 – 5000 mV.
PRES
IN-HG-A
~ - 2”AMBIENT ABSOLUTE
SAMPLE FL
CM
3
/MIN
800 ± 10%
SAMPLE TEMP
ºC
10 – 50 ºC
PHOTO LAMP
ºC
50 ºC ± 1 ºC
O3 GEN TMP*
ºC
48 ºC ± 3 ºC
BOX TEMP
ºC
10 – 50 ºC
SLOPE
1.0
±
.15
OFFSET
PPB
0.0 ± 5.0 PPB
Values are in the Signal I/O
REF_4096_MV
mV
4096mv±2mv and Must be Stable
REF_GND
mV
0± 0.5 and Must be Stable
3. WHAT IS THE SAMPLE FLOW AND SAMPLE PRESSURE WITH THE SAMPLE INLET ON REAR OF MACHINE
CAPPED?
SAMPLE FLOW
-
CM
3
/MIN
SAMPLE PRESSURE
-
IN-HG-A
4. WHAT ARE THE FAILURE SYMPTONS?_________________________________________________________________
_________________________________________________________________________________________________________
_________________________________________________________________________________________________________
5. IF POSSIBLE, PLEASE INCLUDE A PORTION OF A STRIP CHART PERTAINING TO THE PROBLEM. CIRCLE THE
PERTINENT DATA.
6. THANK YOU FOR PROVIDING THIS INFORMATION. YOUR ASSISTANCE ENABLES TELEDYNE API TO RESPOND
FASTER TO THE PROBLEM THAT YOU ARE ENCOUNTERING.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...