APPENDIX A-4: M400E Signal I/O Definitions, Revision .2
Model 400E Instruction Manual
APPENDIX A-4: M400E Signal I/O Definitions, Revision .2
Table A-4: M400E Signal I/O Definitions, Revision C.3
SIGNAL NAME
BIT OR
CHANNEL
NUMBER
DESCRIPTION
Internal inputs, U7, J108, pins 9–16 = bits 0–7, default I/O address 322 hex
0–7
Spare
Internal outputs, U8, J108, pins 1
–
8 = bits 0
–
7, default I/O address 322 hex
0–5
Spare
I2C_RESET
6
1 = reset I2C peripherals
0 = normal
I2C_DRV_RST
7
0 = hardware reset 8584 chip
1 = normal
Control inputs, U11, J1004, pins 1–6 = bits 0–5, default I/O address 321 hex
EXT_ZERO_CAL
0
0 = go into zero calibration
1 = exit zero calibration
EXT_LOW_SPAN_CAL
1
1
0 = go into low span calibration
1 = exit span calibration
EXT_SPAN_CAL
1
2
0 = go into span calibration
1 = exit span calibration
3–5
Spare
6–7
Always
1
Control inputs, U14, J1006, pins 1
–
6 = bits 0
–
5, default I/O address 325 hex
0–5
Spare
6–7
Always
1
Control outputs, U17, J1008, pins 1
–
8 = bits 0
–
7, default I/O address 321 hex
0–7
Spare
Control outputs, U21, J1008, pins 9
–
12 = bits 0
–
3, default I/O address 325 hex
0–3
Spare
Alarm outputs, U21, J1009, pins 1
–
12 = bits 4
–
7, default I/O address 325 hex
ST_SYSTEM_OK2
4
1 = system OK
0 = any alarm condition or in diagnostics mode
5–7
Spare
A status outputs, U24, J1017, pins 1
–
8 = bits 0
–
7, default I/O address 323 hex
ST_SYSTEM_OK
0
0 = system OK
1 = any alarm condition
ST_CONC_VALID
1
0 = conc. valid
1 = hold off or other conditions
ST_HIGH_RANGE
2
0 = high auto-range in use
1 = low auto-range
ST_ZERO_CAL
3
0 = in zero calibration
1 = not in zero
ST_SPAN_CAL
4
0 = in span calibration
1 = not in span
ST_TEMP_ALARM
5
0 = any temperature alarm
1 = all temperatures OK
ST_FLOW_ALARM
6
0 = any flow alarm
1 = all flows OK
ST_PRESS_ALARM
7
0 = any pressure alarm
1 = all pressures OK
16
0
4402
Rev C.3
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...