Model 400E Ozone Analyzer
Instruction Manual
Operating Instructions
P/N 04316 Rev: B
105
To make these adjustments the
AOUT
AutoCal feature must be turned off (see
Section 6.7.3.3) then press:
DIAG AIO
CONC_OUT_1 CALIBRATED: YES
< SET
CAL
EXIT
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED:
NO
< SET
SET>
CAL
EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO
CONC_OUT_1 RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 CALIBRATED: NO
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_1 VOLT–Z : 0 mV
U100 UP10 UP DOWN DN10 D100
ENTR EXIT
DIAG AIO
CONC_OUT_1 :5V, NO CAL
< SET SET>
EDIT
EXIT
Press
SET>
to select the Analog Output
channel to be configured:
DISPLAYED AS = CHANNEL
CONC_OUT_1
=
A1
CONC_OUT_2
=
A2
TEST OUTPUT
=
A4
Then Press
EDIT
to continue
These keys
increment/decrement the
ZERO/SPAN D-to-A converter
output by
100, 10 or 1 counts
respectively.
Continue adjustments until the
voltage measured at the
output of the analyzer and/or
the input of the recording
device matches the value in
the upper right hand corner of
the display to the tolerance
listed in Table 6-11.
The analyzer display
WILL
NOT CHANGE
. Only the
voltage reading of your volt
meter will change.
DIAG AIO
CONC_OUT_1 VOLT–S : 4500 mV
U100 UP10 UP DOWN DN10 D100
ENTR EXIT
EXIT
ignores the
new setting.
ENTR
accepts the
new setting.
DIAG AIO
CONC_OUT_1 REC OFS: 0 mV
< SET
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_1 AUTO CAL: OFF
< SET
SET>
EDIT
EXIT
If AutoCal is
O N
, go to
Section 6.7.3.5.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...