Operating Instructions
Model 400E Ozone Analyzer
Instruction Manual
156
P/N 04316 Rev: B
6.12.6.
Sample Period and Report Period
The iDAS defines two principal time periods by which sample readings are taken
and permanently recorded:
•
Sample Period:
Determines how often iDAS temporarily records a sample
reading of the parameter in volatile memory. The Sample Period is set to
one minute by default and generally cannot be accessed from the standard
iDAS front panel menu, but is available via the instruments communication
ports by using APICOM or the analyzer’s standard serial data protocol.
Sample Period
is only used when the iDAS parameter’s sample mode is set
for
AVG, MIN
or
MAX
.
•
Report Period:
Sets how often the sample readings stored in volatile
memory are processed, (e.g. average, minimum or maximum are calculated)
and the results stored permanently in the instruments Disk-on-Chip as well
as transmitted via the analyzer’s communication ports. The Report Period
may be set from the front panel.
If the
INST
sample mode is selected the instrument stores and reports an
instantaneous reading of the selected parameter at the end of the chosen
Report Period
In
AVG, MIN
or
MAX
sample modes, the settings for the
SAMPLE PERIOD
and
the
REPORT PERIOD
determine the number of data points used each time the
average, minimum or maximum is calculated, stored and reported to the COMM
ports. The actual sample readings are not stored past the end of the of the chosen
REPORT PERIOD
.
Also, the
SAMPLE PERIOD
and
REPORT PERIOD
intervals are synchronized to
the beginning and end of the appropriate interval of the instruments internal clock.
•
If SAMPLE Period were set for one minute the first reading would occur at the
beginning of the next full minute according to the instrument’s internal clock.
•
If the Report period were set for of one hour the first report activity would
occur at the beginning of the next full hour according to the instrument’s
internal clock.
EXAMPLE: Given the above settings, if iDAS were activated at 7:57:35 the first
sample would occur at 7:58 and the first report would be calculated at 8:00
consisting of data points for 7:58. 7:59 and 8:00.
During the next hour (from 8:01 to 9:00) the instrument will take a sample
reading every minute and include 60 sample readings.
When the
STORE NUM. SAMPLES
feature is turned on the instrument will also
store how many sample readings were used for the
AVG, MIN
or
MAX
calculation
but not the readings themselves.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...