Model 400E Ozone Analyzer
Instruction Manual
Operating Instructions
P/N 04316 Rev: B
65
In order to compensate for CPU clocks which run fast or slow, there is a variable to
speed up or slow down the clock by a fixed amount every day. To change this
variable, press:
SAMPLE RANGE = 500.0 PPB O3 =XXX.X
< TST TST > CAL
SETUP
SETUP X.X
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
COMM
VARS
DIAG HALT
EXIT
Exit Returns to
main Sample
Display
SETUP X.X 0 ) DAS_HOLD_OFF=15.0 Minutes
NEXT
JUMP
EDIT PRNT EXIT
Enter Number
of Seconds per
day the clock
gains or loses.
SETUPX.X 1 )PHOTO_LAMP= 52.0 DegC
PREV
NEXT
JUMP
EDIT PRNT EXIT
SETUP X.X
6 ) CLOCK_ADJ=0 Sec/Day
PREV JUMP
EDIT
PRNT EXIT
SETUP X.X
CLOCK_ADJ=
0 Sec/Day
+ 0 0
ENTR
EXIT
SETUP X.X
3 ) CLOCK_ADJ=0 Sec/Day
PREV NEXT JUMP
EDIT PRNT
EXIT
SETUP X.X
ENTER VARS PASS: 818
8 1 8
ENTR
EXIT
Continue to press
NEXT
until …
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...