Operating Instructions
Model 400E Ozone Analyzer
Instruction Manual
60
P/N 04316 Rev: B
To view the TEST Functions press the following Key sequence.
RANGE
STABIL
O3 MEAS
O3 REF
O3 GEN
2
O3 DRIVE
1
PRES
SAMP FL
SAMPLE TEMP
PHOTO LAMP
O3 SCRUB
3
O3 GEN TMP
1
BOX TEMP
SLOPE
TEST
4
OFFSET
TIME
SAMPLE
RANGE = 500.0 PPB O3 =XXX.XX
< TST
TST >
CAL
SETUP
1
Only appears if IZS option is
installed.
2
Only appears if IZS Reference
Sensor option is installed.
3
Only appears if Metal Wool
Scrubber option is installed
4
Only appears if Analog Output
A4 is actively reporting a Test
Function
NOTE
A value of “XXXX” displayed for any of these TEST functions
indicates an OUT OF RANGE reading.
NOTE
Sample Pressure measurements are represented in terms of
ABSOLUTE pressure because this is the least ambiguous method
reporting gas pressure.
Absolute atmospheric pressure is about 29.92 in-Hg-A at sea
level. It decreases about 1 in-Hg per 1000 ft gain in altitude. A
variety of factors such as air conditioning systems, passing
storms, and air temperature, can also cause changes in the
absolute atmospheric pressure.
6.2.3.
Calibration Functions
Pressing the
CAL
key, switches the M400E into calibration mode. In this mode the
user can with the use of calibrated zero, or span gases calibrate the Zero and Full
Span points of the analyzer’s measurement range.
If the instrument includes one of the available Zero/Span Valve options the Sample
Mode display will also include
CALZ
and
CALS
keys. Pressing either of these keys
also put the instrument into Cal Mode. The
CALZ
key is used to initiate a calibration
of the analyzer’s Zero Point. The
CALS
key is used to calibrate the Span point of
the analyzer’s current measurement range. It is recommended that this Span
calibration be performed at 80% of full scale of the analyzer’s Measurement Range.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...