10.3.3.
Optical Bench
Electronically, the Optical Bench is a group of subassemblies that emits UV Light of
the proper wavelength; passes it though a tube filled with Sample Gas; detects the
presence of UV radiation not absorbed by any O
3
present in that Sample Gas and
outputs a voltage signal proportional to the amount of UV detected.
Sample Gas Inlet
Optical Bench
Chassis
Sample Gas
Outlet
Absorption Tube
UV Lamp Heater
& Temperature
Sensor Assy,
UV Lamp
Sample Gas
Temperature
Sensor
UV Lamp
Power Supply
UV Detector
Preamp Assy
UV Detector
Housing
Figure 10-5: Optical Bench Layout – Top View
Photometer UV Lamp
A mercury-vapor UV lamp. This lamp is coated in a material that optically screens
the UV radiation output to remove the O
3
producing 185nm radiation. The majority
of the light emitted is at the 254nm wavelength.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...