Model 400E Instruction Manual
APPENDIX A-3: Warnings and Test Functions, Revision C.3
Table A-3: M400E Test Functions, Revision c.3
TEST Function
Message Text
DESCRIPTION
RANGE
RANGE=500.0 PPB
D/A range in single or auto-range modes.
RANGE1
RANGE1=500.0 PPB
D/A #1 range in dual range mode.
RANGE2
RANGE2=500.0 PPB
D/A #2 range in dual range mode.
STABILITY
STABIL=0.0 PPB
3
Concentration stability (standard deviation based
on setting of STABIL_FREQ and
STABIL_SAMPLES).
PHOTOMEAS
O3 MEAS=2993.8 MV
Photometer detector measure reading.
PHOTOREF
O3 REF=3000.0 MV
Photometer detector reference reading.
O3GENREF
O3 GEN=4250.0 MV
O3 generator reference detector reading.
O3GENDRIVE
O3 DRIVE=0.0 MV
O3 generator lamp drive output.
PHOTOPOWER
PHOTO POWER=4500.0 MV
Photometer lamp drive output.
SAMPPRESS
PRES=29.9 IN-HG-A
Sample pressure.
SAMPFLOW
SAMP FL=700 CC/M
Sample flow rate.
SAMPTEMP
SAMPLE TEMP=31.2 °C
Sample temperature.
PHOTOLTEMP
PHOTO LAMP=52.3 °C
Photometer lamp temperature.
O3SCRUBTEMP
O3 SCRUB=110.2 °C
O3 scrubber temperature.
O3GENTEMP
O3 GEN TMP=48.5 °C
O3 generator lamp temperature.
BOXTEMP
BOX TEMP=31.2 °C
Internal chassis temperature.
SLOPE
SLOPE=1.000
Slope for current range, computed during
zero/span calibration.
OFFSET
OFFSET=0.0 PPB
Offset for current range, computed during
zero/span calibration.
O3
O3=191.6 PPB
O3 concentration for current range.
TESTCHAN
TEST=2753.9 MV
Value output to TEST_OUTPUT analog output,
selected with TEST_CHAN_ID variable.
CLOCKTIME
TIME=14:48:01
Current instrument time of day clock.
0
4402
Rev C.3
15
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...