Operating Instructions
Model 400E Ozone Analyzer
Instruction Manual
80
P/N 04316 Rev: B
The Test Functions available to be reported are:
Table 6-5: Test Functions available to the A4 Analog Output Channel
Test Channel
Zero
Full Scale
NONE
Test Channel is turned off
PHOTO MEAS
0 mV
5000 mV*
PHOTO REF
0 mV
5000 mV*
O3 GEN REF
0 mV
5000 mV*
SAMPLE PRESS
0 "Hg
40 "Hg
SAMPLE FLOW
0 cc/m
1000 cc/m
SAMPLE TEMP
0 C
°
70 C
°
PHOTO LAMP TEMP
0 C
°
70 C
°
O3 SCRUB TEMP
0 C
°
70 C
°
03 LAMP TEMP
0 mV
5000 mV
CHASSIS TEMP
0 C
°
70 C
°
* This refers to the internal voltage level of the function NOT the output
signal level of the Test channel itself.
Once a function is selected, the instrument not only begins to output a signal on the
A4
Analog Output, but adds
TEST
to the list of Test Functions viewable via the
Front Panel Display.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...