MAINTENANCE SCHEDULE & PROCEDURES
Model 400E Ozone Analyzer
Instruction Manual
214
P/N 04316 Rev: B
9.3.6.
Flow Calibration
Once an accurate measurement has been recorded by the method described above,
adjust the analyzer’s internal flow sensors by pressing:
SAMPLE RANGE = 500.0 PPB O3 =XXX.X
< TST TST > CAL
S
ETUP
SETUP X.X
CFG ACAL DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X
COMM VARS
DIAG
HALT
EXIT
Exit at Any
Time to Return
to main the
SETUP Menu
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Exit returns
to the
previous menu
SETUP X.X
ENTER DIAG PASS: 818
8 1 8
ENTR
EXIT
Repeat Pressing
NEXT
until . .
.
DIAG
FLOW CALIBRATION
PREV NEXT
ENTR EXIT
DIAG FCAL
ACTUAL FLOW: 780 CC / M
0 7 8 0 ENTR EXIT
E
NTR
accepts the
new value and
returns to the
previous menu
E
XIT
ignores the
new value and
returns to the
previous menu
Toggle these keys until
the displayed flow rate
equals the flow rate being
measured by the
independent flow meter.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...