Model 400E Ozone Analyzer
Instruction Manual
THEORY OF OPERATION
P/N 04316 Rev: B
217
10.
THEORY OF OPERATION
The Model 400E Ozone Analyzer is a microprocessor controlled analyzer that
determines the concentration of Ozone (O
3
) in a sample gas drawn through the
instrument. It requires that sample and calibration gasses be supplied at ambient
atmospheric pressure in order to establish a stable gas flow through the Absorption
Tube where the gas’ ability to absorb ultraviolet (UV) radiation of a certain
wavelength (in this case 254 nm) is measured.
Calibration of the instrument is performed in software and does not require physical
adjustments to the instrument. During calibration the microprocessor measures the
current state of the UV Sensor output and various other physical parameters of the
instrument and stores them in memory.
The microprocessor uses these calibration values, the UV absorption measurements
made on the Sample Gas in the Absorption Tube along with data regarding the
current temperature and pressure of the gas to calculate a final O
3
concentration.
This concentration value and the original information from which it was calculated
are stored in one of the unit’s Internal Data Acquisition System (iDAS - see
Sections 6.11 and 6.12) as well as reported to the user via a Front Panel Display or
a variety of digital and analog signal outputs.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...