Model 400E Ozone Analyzer
Instruction Manual
THEORY OF OPERATION
P/N 04316 Rev: B
225
10.2.3.
Particulate Filter
The Model 400E Ozone Analyzer comes equipped with a 47 mm diameter Teflon
particulate filter with a 5 micron pore size.
The filter is accessible through the front panel, which folds down to allow access,
and should be changed according to the suggested maintenance schedule described
in Table 9-1.
10.2.4.
Zero Span Gas Supply Options
Several options can be purchased for the analyzer which allow the user to more easily
supply and manipulate Zero Air and Span Gas.
The available options are:
Option
Description
Delivery Conditions
50
Zero/Span Valve with
Sample/Cal valve.
Allows the use of calibration gasses, such as Zero
Air and Span Gas, from sources external to the
instrument. Gasses must be supplied at ambient
atmospheric pressure.
During instrument calibration Zero Air and Span
Gas flow is regulated internally.
51
Internal Zero / Span
Option (IZS) with
Sample/Cal valve
Built in O
3
generator turns on and off to Internally
generates zero Air and Span Gas. Sample/Cal valve
switches between IZS supplied gas and gas from
the instrument’s Sample Inlet.
May be used for performing Calibration
CHECKS
ONLY
.
IZS option is not intended for use during instrument
Calibration.
For more information of these options see Section 5.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...