6.6.2.
Analog Output Calibration Mode
The analog outputs can be calibrated automatically or manually. In its default
mode, the instrument is configured for automatic calibration of all channels. Manual
calibration should be used for the 0.1V range or in cases where the outputs must be
closely matched to the characteristics of the recording device. Outputs configured
for automatic calibration can be calibrated as a group or individually. Calibration of
the analog outputs needs to be carried out on first startup of the analyzer
(performed in the factory as part of the configuration process) or whenever re-
calibration is required.
6.6.2.1.
Automatic Analog Output Calibration
To calibrate the outputs as a group, activate the analog i/o configuration menu,
then press:
DIAG AIO
AUTO
CALIBRATING CONC_OUT_1
AUTO
CALIBRATING CONC_OUT_2
AUTO CALIBRATING TEST_OUTPUT
Exit at any time
to return to the
main DIAG
menu
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED:
NO
< SET
SET>
CAL
EXIT
If AutoCal has been
turned off for any
channel, the message
for that channel will be
similar to:
NOT AUTO CAL
CONC_OUT_1
Exit to return to
the I/O
configuration
menu
STARTING FROM DIAGNOSTIC MENU
(see Section 6.6)
If any of the channels have
not been calibrated this
message will read
NO
.
DIAG AIO AOUTS CALIBRATED:
YES
< SET
SET> CAL
EXIT
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...