Operating Instructions
Model 400E Ozone Analyzer
Instruction Manual
154
P/N 04316 Rev: B
6.12.4.
Adding or Deleting Data Parameters
To add or delet a Data Parameter to a Data Channel, press:
Edit Data Parameter Menu
Edit Data Channel Menu
SETUP X.X
0)
CONC: ATIMER, 2, 900
PREV NEXT INS DEL
EDIT
PRNT
EXIT
FROM THE DATA CHANNEL PASSWORD MENU
Exits to the Main
Data Acquisition
Menu
SETUP X.X
NAME:CONC
<SET
SET>
EDIT
EXIT
SETUP X.X
PARAMETERS:2
<SET SET>
EDIT
EXIT
SETUP X.X
0) PARAM=DETREP, MODE=INST
PREV NEXT
INS DEL EDIT EXIT
Press
SET>
key until…
SETUP X.X
EDIT PARAMS (DELETE DATA)
YES NO
NO
returns to
the previous
menu.
Moves the
display up &
down the list of
Parameters
Inserts a new Data
Parameter into the list
BEFORE the one
currently being displayed
Deletes The Data Parameter
currently being displayed
AND ALL DATA
RECORDED FOR THAT
PARAMETER TO DATE
User to configure
the FUNCTIONS
for this Data
Parameter.
Exits to the Main
Data Acquisition
Menu
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...