11.4.1.3.
IZS Ozone Generator Temperature (Optional)
There are three possible causes for the Ozone Generator temperature to have
failed.
1.
The O3 Gen heater has failed. Check the resistance between pins 5 and 6 on
the six-pin connector adjacent to the UV Lamp on the O3 Generator. It
should be approximately 5 Ohms.
2.
Assuming that the I
2
C bus is working and that there is no other failure with
the Relay board, the FET Driver on the Relay Board (see Figure 10-2) may
have failed. Using the
O3_GEN_HEATER
parameter under the
SIGNAL I/O
function of the
DIAG
menu, as described above, turn on and off the UV
Lamp Heater. Check the DC voltage present between pin 1 and 2 on J14 of
the Relay Board.
If the FET Driver has failed there should be no change in the voltage across
pins 1 and 2.
3.
If the FET Driver checks out OK, the thermistor temperature sensor in the
lamp assembly may have failed. Unplug the connector to the Ozone
Generator Heater/Thermistor PCB, and measure the resistance of the
thermistor between pins 5 and 6 of the 6 pin connector.
11.5.
Subsystem Checkout
The preceding sections of this manual discussed a variety of methods for identifying
possible sources of failures or performance problems within the analyzer. In most
cases this included a list of possible causes. This section describes how to
determine individually determine if a certain component or subsystem is actually
the cause of the problem being investigated.
11.5.1.
AC Mains Configuration
The analyzer is correctly configured for the AC mains voltage in use if:
•
The Sample Pump is running.
If incorrect power is suspected, check that the correct voltage and frequency is
present at the line input on the rear panel.
•
If the unit is set for 230 VAC and is plugged into 115VAC, or 100VAC the sample
pump will not start.
•
If the unit is set for 115 or 100 VAC and is plugged into a 230 VAC circuit, the
circuit breaker built into the ON/OFF Switch on the Front Panel will trip to the
OFF position immediately after power is switched on.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...