
TROUBLESHOOTING & REPAIR PROCEDURES
Model 400E Ozone Analyzer
Instruction Manual
262
P/N 04316 Rev: B
11.5.6.3.
Status Outputs
The procedure below can be used to test the Status outputs:
1.
Connect a jumper between the “D“ pin and the “
s
” pin on the status output
connector.
2.
Connect a 1000 ohm resistor between the +5V and the pin for the status output
that is being tested.
3.
Connect a voltmeter between the “D“ pin and the pin of the output being tested
(see table below).
4.
Under the
DIAG
à
SIGNAL I/O
menu (see Section 11.1.3), scroll through the
inputs and outputs until you get to the output in question. Alternately turn on
and off the output noting the voltage on the Voltmeter, it should vary between
0 volts for ON and 5 volts for OFF.
Table 11-7: Status Outputs Check
Pin
(LEFT TO RIGHT)
Status
1
SYSTEM OK
2
CONC VALID
3
HIGH RANGE
4
ZERO CAL
5
SPAN CAL
6
DIAG MODE
7
SPARE
8
SPARE
11.5.6.4.
Control Inputs – Remote Zero, Span
The control input bits can be tested by the following procedure:
1.
Connect a jumper from the “+” pin on the Status connector to the “U” on the
Control In connector.
2.
Connect a second jumper from the Digital Ground pin on the Control In
connector to the A pin. The instrument should switch from
SAMPLE
mode to
ZERO CAL R
mode.
3.
Connect a second jumper from the Digital Ground pin on the Control In
connector to the B pin. The instrument should switch from
SAMPLE
mode to
SPAN CAL R
mode.
In each case, the M400E should return to SAMPLE mode when the jumper is
removed.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...