Model 400E Ozone Analyzer
Instruction Manual
Operating Instructions
P/N 04316 Rev: B
127
The card has four LEDs that are visible on the rear panel of the analyzer, indicating
its current operating status.
Table 2-7: Ethernet Status Indicators
LED
FUNCTION
LNK (green)
ON when connection to the LAN is valid.
ACT (yellow)
Flickers on any activity on the LAN.
TxD (green)
Flickers when the RS-232 port is
transmitting data.
RxD (yellow)
Flickers when the RS-232 port is
receiving data.
This option can be installed in conjunction with the RS-232 multidrop (option 62)
allowing the instrument to communicate on both types of networks simultaneously.
6.9.8.1.
Ethernet Card COM2 Communication Modes and
Baud Rate
The COM2 baud rate and communication modes for the COM2 are automatically set
when the Ethernet option is enabled. The baud rate is automatically set at 115 200
kBaud.
6.9.8.2.
Configuring the Ethernet Interface Option using
DHCP
The Ethernet option uses Dynamic Host Configuration Protocol (DHCP) to
automatically configure its interface with your LAN. This requires your network
servers also be running DHCP. The analyzer will do this the first time you turn the
instrument on after it has been physically connected to your network. Once the
instrument is connected and turned on it will appear as an active device on your
network without any extra set up steps or lengthy procedures.
Should you need to, the following Ethernet configuration properties are viewable via
the analyzer’s front panel.
Table 2-8: LAN/Internet Configuration Properties
PROPERTY
DEFAULT STATE
DESCRIPTION
DHCP
STATUS
On
Editable
This displays whether the DHCP is
turned ON or OFF.
INSTRUMENT
IP ADDRESS
Configured
by DHCP
EDIT
key
disabled
when DHCP
is
ON
This string of four packets of 1 to 3
numbers each (e.g. 192.168.76.55.) is
the address of the analyzer itself.
Содержание 400E
Страница 12: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 12 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 16: ...M400E DOCUMENTATION Model 400E Ozone Analyzer Instruction Manual 16 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 46: ...Frequently Asked Questions Model 400E Ozone Analyzer Instruction Manual 46 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 184: ...CALIBRATION PROCEDURES Model 400E Ozone Analyzer Instruction Manual 184 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 204: ...EPA Protocol Calibration Model 400E Ozone Analyzer Instruction Manual 204 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 216: ...MAINTENANCE SCHEDULE PROCEDURES Model 400E Ozone Analyzer Instruction Manual 216 P N 04316 Rev B INTENTIONALLY BLANK ...
Страница 294: ...Model 400E Ozone Analyzer Instruction Manual 2 04404 Rev B INTENTIONALLY BLANK ...